This document provides a recommended room layout and equipment placement for an Allura FD20 Ceiling angiography suite. Key equipment such as the generator, peripheral cabinet, and mains cabinet are labeled and their weights and heat outputs listed. Environmental requirements specify maintaining room temperature between 72°F +/- 10°F and humidity between 45% +/- 15%. Power requirements include a 3-phase, 3 or 4-wire configuration with 380-480VAC and a 225KVA branch circuit.