Can I read Optimizing Harmonics Measurement with Samplers on EtoBox?
Optimizing Harmonics Measurement with Samplers by Gokulakrishnan is a document available to read on EtoBox.
What is Optimizing Harmonics Measurement with Samplers about?
1. The document discusses methods for accurately measuring harmonic distortion using a sampler. 2. It finds that applying attenuation of around 17dB to lower the input signal amplitude to around -30dBm allows the sampler to more accurately measure harmonic distortion. 3. Verification measurements comparing the sampler to a spectrum analyzer show good agreement between the two instruments when using the recommended 17dB attenuation, validating that the sampler can accurately measure harmonic distortion u
- Author
- Gokulakrishnan
- Language
- EN