Skip to content

Opening book details…

Can I read A data adaptive reprojection technique for MR angiography on EtoBox?

A data adaptive reprojection technique for MR angiography by F. R. Korosec; D. M. Weber; C. A. Mistretta; P. A. Turski; M. A. Bernsteins is a scholarly article available to read on EtoBox.

What is A data adaptive reprojection technique for MR angiography about?

## Abstract Inability to detect vessel overlap and vascular loops can compromise the interpretation of magnetic resonance angiograms. A data‐adaptive ray tracing (DART) technique was developed to produce the appropriate variations in signal intensity at points of vessel overlap in order to simulate the standard angiographic representation of vessels. In this technique a threshold is utilized to identify vessels in the image slices composing a 3D angiographic data set. A mask, which defines regions slightly larger than the vessel boundaries, is obtained by blurring the vessel information surviving the initial threshold. This mask is converted to binary form prior to multiplication by the original angiographic data set. Following application of an additional threshold to the masked data, line integrals through the regions defined by the mask are performed to obtain an angiographic signal proportional to the integrated vessel signal as in conventional angiography. This integrated reprojection is then uniquely combined with a maximum intensity pixel (MIP) reprojection to produce the final DART image. The application of the DART technique to 2D time‐of‐flight and 3D phase‐contrast angio

Author
F. R. Korosec; D. M. Weber; C. A. Mistretta; P. A. Turski; M. A. Bernsteins
Publisher
John Wiley and Sons; Wiley (John Wiley & Sons); John Wiley & Sons Inc.; Wiley (ISSN 0740-3194)
Published
1992
Language
EN

More by F. R. Korosec; D. M. Weber; C. A. Mistretta; P. A. Turski; M. A. Bernsteins

Browse all works by F. R. Korosec; D. M. Weber; C. A. Mistretta; P. A. Turski; M. A. Bernsteins