Optimizing Gas Chromatography for Methyl Esters by batrisyia hazirah is a document available to read on EtoBox.
This experiment investigated the optimization of gas chromatography (GC) separation of compounds in a methyl ester mixture. Two factors were examined: carrier gas flow rate and column temperature. At a flow rate of 70 cm/s and temperature of 210°C, the methyl ester compounds achieved baseline separation as measured by a resolution over 1.5, while also having the shortest retention times, making it the optimal condition. Testing different flow rates of 30, 50, and 70 cm/s at 210°C and temperatures of 170, 19
- Author
- batrisyia hazirah
- Language
- EN