X-ray tubes are measured in heat units, which are calculated using formulas that factor in voltage, current, and exposure time. Higher techniques produce more heat units. For single phase generators, a sample technique of 85kVp at 4mA for 2 seconds equals 3570 heat units, which is within safe limits. However, a three phase generator technique of 155kVp at 350mA for 2 seconds equals 151,900 heat units, which is too high and risks overheating the tube. Similarly, a high frequency generator technique of 65kVp