The Karl Fischer (KF) Oven Method permits the accurate, precise and straightforward determination of the water content of pharmaceuticals that are not amenable to direct KF titration because they either undergo detrimental side reactions with the KF reagent or release their water only slowly or at high temperatures. Water determinations carried out with several pharmaceuticals demonstrate that the values obtained with the KF Oven Method all lie within the ranges specified in the European Pharmac