Can I read A Microelectrode Flow Amperometric Detector for Water-soluble Polymers on EtoBox?

A Microelectrode Flow Amperometric Detector for Water-soluble Polymers by Victoria A. Vicente-Beckett; Min-Kuang Liu; J.Calvin Giddings; Stanley Pons; Marcus N. Myers is a Chemistry article available to read on EtoBox.

What is A Microelectrode Flow Amperometric Detector for Water-soluble Polymers about?

A single electrode assembly, consisting of a platinum wire (20-or 25pm diameter) microelectrode and a palladium coating (as auxiliary/pseudo-reference electrode) on the glass tubing housing the microelectrode, has been utilized as a flow amperometric sensor for water-soluble polymers. An ordinary 3-V dry cell was used as voltage source. Amounts of injected polymer were in the range 0.01-260 ,ug. The detector performance was superior to that of a differential refractive index detector in terms of limits of detection (sub-pg levels of polymer), baseline stability, and reproducibility of analytical signal.

Who reads A Microelectrode Flow Amperometric Detector for Water-soluble Polymers?

It is typically read by researchers, students, and practitioners in Chemistry.