Ellman EMC & Dento Test Procedure by SBA KALIBRASI is a document available to read on EtoBox.

This document provides a test procedure for measuring electromagnetic compatibility (EMC) and power output of electrosurgical units. It involves: 1) Calibrating the oscilloscope daily before measurements. 2) Recalling setup parameters and connecting the load and oscilloscope probes. 3) Testing each cutting, cutting/coagulation, coagulation, and fulguration mode from levels 1 to high and verifying power output falls within specified limits. 4) Recording results in a quality record if a medical unit.