This document validates the MUSE wearable EEG headset for frequency domain analysis and measuring frontal alpha asymmetry (FAA). It compares spectral measures from the low-cost 4-channel MUSE headset to a research-grade 64-channel EEG system. Results show the MUSE can examine power spectral density across frequencies and measure FAA and individual alpha frequency, with good internal consistency. However, estimating asymmetry on the IAF did not provide advantages over the traditional 8-13 Hz range. These fin