Electrosurgical Equipment Safety Study by Maulana Bastian is a document available to read on EtoBox.

This study evaluates the performance and safety of electrosurgical equipment in six hospitals in Iran, highlighting the critical need for improved regulations and training for medical staff. Results indicated that high-frequency leakage currents were higher in ground-referenced generators compared to isolated ones, with only eight out of twenty units delivering acceptable output values. The findings emphasize the importance of regular performance verifications to enhance patient safety in surgical environme

Author
Maulana Bastian
Language
EN