This research investigates the containment efficiency of laboratory chemical hoods at different face velocities, specifically comparing occupied hoods at 100 fpm to unoccupied hoods at 60 fpm. The results indicate that unoccupied hoods can maintain containment as effectively, if not better, than occupied hoods under ideal conditions, suggesting potential for reduced energy consumption. Further testing is recommended to assess performance in real-world scenarios with additional challenges present.