Can I read A Comparison of Single- and Dual-beam Methods for Maximum Velocity Estimation on EtoBox?
A Comparison of Single- and Dual-beam Methods for Maximum Velocity Estimation by P.R Hoskins is a Medicine article available to read on EtoBox.
What is A Comparison of Single- and Dual-beam Methods for Maximum Velocity Estimation about?
The purpose of this study was to compare the precision and accuracy of maximum velocity estimation when the target direction is known (a string phantom), and when the target direction is unknown (a flow model of arterial stenosis with stenoses of 0 -80% by area). Maximum velocity was estimated using single-and dual-beam methods. A linear-array system was used to acquire Doppler spectra from a single-beam direction. The same array was used for sequential acquisition of Doppler spectra from 2 beam directions; the velocity estimates from these were then compounded in a vector manner. The variation of estimated maximum velocity with beam-string angle over the range 40 -80°was 27% for conventional Doppler, 2.6% for angle correction from the edge of the array and 1.6% for the vector Doppler. In the stenosis model, for the single-beam methods, the highest frequency shift was obtained just prior to the point of minimum lumen. At this location, the variation with beam-vessel angle over the range 40 -80°was 35% for conventional Doppler, 7.4% for the correction factor method and 6.9% for correction from the edge of the array. For the vector method, the maximum velocity is obtained from within
Who reads A Comparison of Single- and Dual-beam Methods for Maximum Velocity Estimation?
It is typically read by researchers, students, and practitioners in Medicine.
- Author
- P.R Hoskins
- Publisher
- Elsevier Science; Elsevier ; Elsevier BV (ISSN 0301-5629)
- Published
- 1999
- Language
- EN
- Field