About this document
Laser Tracker Calibration Procedures by cmm5477 is a document available to read on EtoBox.
This document summarizes the laser tracker calibration procedure at the Central Office of Measures in Poland. It describes setting up an ASME B89.4.19 laser tracker verification facility to assess instrument performance. Tests performed include laser wavelength calibration, environmental sensor calibration, ranging tests over 50 meters, and volumetric length measurement tests using a 2.31 meter carbon composite reference artifact. Example test results are presented, showing the laser tracker
- Author
- cmm5477
- Language
- EN