Internal Standards for ICP-MS Analysis by Darma A. Wardhana is a document available to read on EtoBox.
The document discusses using a self-organizing map to select optimal internal standards for determining certain elements like Cd, Pb, Sn, and platinum group elements in fish and mollusk samples using ICP-MS. It evaluates different internal standards and analyte/internal standard pairs, finding the best pairs to be 114Cd+/74Ge+, 195Pt+/74Ge+, and 208Pb+/74Ge+. Recoveries for analytes using the selected internal standards or without ranged from 98 to 101%.
- Author
- Darma A. Wardhana
- Language
- EN