Instruments, gloves, and glass syringes can be processed for reuse by: Wearing gloves and soaking items in a 0.5% chlorine solution for 10 minutes. Thoroughly washing and rinsing the instruments. Acceptable sterilization methods include autoclaving at 121°C for 20 minutes, hot air oven at 160°C for 1 hour, or boiling and steaming items for 20 minutes after the water boils. Chemical sterilization methods include soaking in 2% glutaraldehyde solution for 10-24 hours. A 0.5% chlorine solution for decontaminati