The DHEA RIA assay, catalogue number DSL8900, has been recalibrated from kit lot 101105, effective November 15, 2010. The new calibration aligns with expected normal sample values, resulting in patient sample values being approximately 70% lower than the previous calibration. For further details, contact your local representative.