2998 PDA Detector Initialization Issues by islampes9 is a document available to read on EtoBox.
This document provides troubleshooting steps for a 2998 PDA Detector initialization failure issue where the lamp light is red. The cause is determined to be a dry flow cell. The recommended fix is to flush the detector with 100% methanol at 2.0 mL/min for at least 15 minutes before powering on to allow initialization and verification to complete successfully. Additional tips note that rebooting the detector during flow or resetting from the console may also resolve the issue if flushing does not work.