Skip to content

Opening book details…

About this document

Nitric Oxide Impact on ASTM D5453 Testing by Jacek is a document available to read on EtoBox.

The document discusses the ASTM D5453 standard for determining nitric oxide interference in UV-F sulfur determination. It describes a test procedure involving measuring sulfur concentrations from samples with increasing nitrogen levels to check for interference. TE Instruments performed this test using a butyl sulfide standard and found that correction for nitrogen interference was not required, as sulfur measurements stayed within the repeatability limit of 1.0 mg/kg even at nitrogen levels up to 300 mg/kg

Author
Jacek
Language
EN