Omron EVOLV Blood Pressure Monitor Validation by Kokou Gbegan is a document available to read on EtoBox.

The study validated the Omron EVOLV upper arm blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. 33 subjects were recruited from a general population ranging in age from 26 to 91 years old. Measurements from the device and two observers were within 5 mmHg for 73% of systolic and 81% of diastolic readings. The device met all validation requirements and is recommended for personal use in a general population.

Author
Kokou Gbegan
Language
EN