Skip to content

Opening book details…

Can I read Fano Factor and Noise in HPGe Detectors on EtoBox?

Fano Factor and Noise in HPGe Detectors by Claretta is a document available to read on EtoBox.

What is Fano Factor and Noise in HPGe Detectors about?

This document describes an experiment to determine the Fano factor and pre-amplifier noise of a HPGe detector. A gamma-ray detector and standard solution containing various radioactive isotopes were used to collect energy and resolution data pairs. Graphs of resolution squared versus energy were plotted, and the slope and intercept were used to calculate the Fano factor and pre-amplifier noise, respectively. The average Fano factor obtained was 0.123±0.003, consistent with the typical value of 0.13 for semi

Author
Claretta
Language
EN