Opening book details…
Can I read Biopsy needles and devices: Assessment of ferromagnetism and artifacts during exposure to a 1.5-T MR system on EtoBox?
Biopsy needles and devices: Assessment of ferromagnetism and artifacts during exposure to a 1.5-T MR system by Marleigh A. Moscatel; Frank G. Shellock; Stacy M. Morisoli is a Medicine article available to read on EtoBox.
What is Biopsy needles and devices: Assessment of ferromagnetism and artifacts during exposure to a 1.5-T MR system about?
## Forty -eight biopsy devices (needles, stylets, guide wires. biopsy guns. etc) were evaluated for ferromagnetism (n = 48) and artifacts (n = 43) in a 1.5-T magnetic resonance (MR) system, with previously deflve of the 48 devices were found to be ferromagnetic. Artifacts varied, depending on the type and size of the component material(s) of the device. Most of the commercially available biopsy devices are not useful opsy procedures because of the presence of ferromagnetism and associated artifacts. scribed techniques. Fortyfor M R imaghg-guided bil n d u t-: Arufact -Blopsles. technology \* Magnetlr resonance (MR) guldance \* Safety
Who reads Biopsy needles and devices: Assessment of ferromagnetism and artifacts during exposure to a 1.5-T MR system?
It is typically read by researchers, students, and practitioners in Medicine.
- Author
- Marleigh A. Moscatel; Frank G. Shellock; Stacy M. Morisoli
- Publisher
- John Wiley and Sons; Wiley (John Wiley & Sons); John Wiley & Sons Inc.; Wiley (ISSN 1053-1807)
- Published
- 1995
- Language
- EN
- Field
- Medicine (Health Sciences)