Tn-0132 Ozone Sensorpolarizationcalib by Mohamed Kamal is a document available to read on EtoBox.
What is Tn-0132 Ozone Sensorpolarizationcalib about?
The document discusses recommendations for polarizing and calibrating ozone sensors used in pharmaceutical water ozonation systems. It recommends polarizing the sensors for at least 2 hours at >0.05 ppm ozone or 30 minutes at >0.5 ppm ozone before calibration. Span calibration should be done at >0.25 ppm ozone during system sanitization. Zero calibration removes the sensor from flow and exposes it to air for 10 minutes.