Opening book details…
Can I read A CT using longitudinally detected ESR (LODESR-CT) of intraperitoneally injected nitroxide radical in a rat's head on EtoBox?
A CT using longitudinally detected ESR (LODESR-CT) of intraperitoneally injected nitroxide radical in a rat's head by Hidekatsu Yokoyama; Toshiyuki Sato; Nobuaki Tsuchihashi; Tateaki Ogata; Hiroaki Ohya-Nishiguchi; Hitoshi Kamada is a Medicine article available to read on EtoBox.
What is A CT using longitudinally detected ESR (LODESR-CT) of intraperitoneally injected nitroxide radical in a rat's head about?
We have developed an in vivo longitudinally detected ESR (LODESR) imaging system operating at 700 MHz based on a loop-gap resonator and a pair of saddle-type pickup coils. A good linear relationship between the LODESR signal intensity and the applied power in a range up to 15.8 W was obtained. The detection of LODESR signals was barely affected by variations in the resonant properties. The characteristic of LODESR is suitable for in vivo examination. Using this system, we succeeded in obtaining LODESR-CT images of the head region of a rat after the intraperitoneal injection of a nitroxide radical.
Who reads A CT using longitudinally detected ESR (LODESR-CT) of intraperitoneally injected nitroxide radical in a rat's head?
It is typically read by researchers, students, and practitioners in Medicine.
- Author
- Hidekatsu Yokoyama; Toshiyuki Sato; Nobuaki Tsuchihashi; Tateaki Ogata; Hiroaki Ohya-Nishiguchi; Hitoshi Kamada
- Publisher
- Elsevier Science; Elsevier ; Elsevier BV (ISSN 0730-725X)
- Published
- 1997
- Language
- EN
- Field
- Medicine (Health Sciences)