1) Visually inspect the driver assemblies for signs of arcing or damage as blown drivers often indicate an arcing x-ray tube, which may have also damaged the high-voltage transformer tank. 2) Replacing just the drivers and tank may not solve the problem if the x-ray tube is arcing, so installing a new known good tube is recommended to avoid further damage. 3) Other potential causes of kV errors include a bad kV control board, though drivers/tube issues are more common, so inspecting those should be the