Infrared Thermometer Calibration Guide by marcelo.lackner is a document available to read on EtoBox.
This document describes testing the suitability of infrared cavity radiators used to verify infrared thermometers. It specifies comparing a calibration radiator to a reference radiator under laboratory conditions. The calibration radiator must not deviate from the reference by more than 0.07°C at three temperatures and have a mean deviation of no more than 0.12°C to be suitable.