Artifacts and signal loss due to flow in the presence of B~o~ inhomogeneity by Maria Drangova; Norbert J. Pelc is a Medicine article available to read on EtoBox.

What is Artifacts and signal loss due to flow in the presence of B~o~ inhomogeneity about?

## Abstract An __in vitro__ study was performed to investigate the effects of B~o~ inhomogeneity on magnetic resonance images of flow. Controlled inhomogeneity gradients (G~1~) were applied and the magnitude of the artifacts produced was quantified for different echo delay times (__TE__). Both steady and pulsatile flows were examined. In the presence of an inhomogeneity gradient, signal loss is apparent if the flow is pulsatile and/or if the slice thickness is large. The signal loss increases with increasing __TE__ and G~1~. With pulsatile flow, ghosting artifacts are also generated. These increase in intensity with increasing __TE__ and G~1~. __In vivo__, field inhomogeneity due to susceptibility variations is large enough to produce these effects. Representative time‐of‐flight images obtained of a normal volunteer with two different TEs demonstrate the effect __in vivo__. Flow‐related signal loss and artifacts, therefore, increase with increasing TE independent of the moments of the applied gradients.

Who reads Artifacts and signal loss due to flow in the presence of B~o~ inhomogeneity?

It is typically read by researchers, students, and practitioners in Medicine.

Author
Maria Drangova; Norbert J. Pelc
Publisher
John Wiley and Sons; Wiley (John Wiley & Sons); John Wiley & Sons Inc.; Wiley (ISSN 0740-3194)
Published
1996
Language
EN
Field
Medicine (Health Sciences)

More by Maria Drangova; Norbert J. Pelc

Browse all works by Maria Drangova; Norbert J. Pelc